PERFORMANCE

682XXB/683XXB

VERIFICATION

POWER LEVELS

Table 3-2.682XXB/683XXB Maximum Leveled Output Power (2 of 2)

 

68XXXB

Frequency

Max Leveled

Max Leveled

 

 

Output Power

 

 

Model

 

(GHz)

Output Power

 

 

 

w/Step Attenuator

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.5

– 2.2 GHz

+11.0 dBm

 

 

 

 

2.2 – 20.0 GHz

+10.0 dBm

 

 

 

68X95B

20.0

– 40.0 GHz

+2.5 dBm

Not Available

 

 

 

40.0

– 50.0 GHz

0.0 dBm

 

 

 

 

50.0 – 65.0 GH

–2.0 dBm

 

 

 

 

 

 

 

 

 

 

0.01 – 2.0 GHz

+12.0 dBm

 

 

 

 

2.0 – 20.0 GHz

+10.0 dBm

 

 

 

68X97B

20.0

– 40.0 GHz

+2.5 dBm

Not Available

 

 

 

40.0

– 50.0 GHz

0.0 dBm

 

 

 

 

50.0

– 65.0 GHz

–2.0 dBm

 

 

 

 

 

 

 

 

With Option 15B (High Power) Installed

 

 

 

 

 

 

 

 

68X37B

2.0 – 20.0 GHz

+17.0 dBm

+15.0 dBm

 

 

 

 

 

 

 

 

 

68X45B

0.5

– 2.2 GHz

+13.0 dBm

+11.0 dBm

 

 

2.2 – 20.0 GHz

+17.0 dBm

+15.0 dBm

 

 

 

 

 

 

 

 

 

 

 

68X47B

0.01 – 2.0 GHz

+13.0 dBm

+11.0 dBm

 

 

2.0 – 20.0 GHz

+17.0 dBm

+15.0 dBm

 

 

 

 

 

 

 

 

 

 

 

68X53B

2.0 – 20.0 GHz

+13.0 dBm

+11.0 dBm

 

 

20.0

– 26.5 GHz

+10.0 dBm

+7.5 dBm

 

 

 

 

 

 

 

 

 

 

 

 

 

0.5

– 2.2 GHz

+13.0 dBm

+11.0 dBm

 

 

68X55B

2.2 – 20.0 GHz

+13.0 dBm

+11.0 dBm

 

 

 

20.0

– 26.5 GHz

+10.0 dBm

+7.5 dBm

 

 

 

 

 

 

 

 

 

0.01 – 2.0 GHz

+13.0 dBm

+11.0 dBm

 

 

68X59B

2.0 – 20.0 GHz

+13.0 dBm

+11.0 dBm

 

 

 

20.0

– 26.5 GHz

+10.0 dBm

+7.5 dBm

 

 

 

 

 

 

 

 

68X63B

2.0 – 20.0 GHz

+13.0 dBm

+11.0 dBm

 

 

20.0

– 40.0 GHz

+6.0 dBm

+3.0 dBm

 

 

 

 

 

 

 

 

 

 

 

 

 

0.5

– 2.2 GHz

+13.0 dBm

+11.0 dBm

 

 

68X65B

2.2 – 20.0 GHz

+13.0 dBm

+11.0 dBm

 

 

 

20.0

– 40.0 GHz

+6.0 dBm

+3.0 dBm

 

 

 

 

 

 

 

 

 

0.01 – 2.0 GHz

+13.0 dBm

+11.0 dBm

 

 

68X69B

2.0 – 20.0 GHz

+13.0 dBm

+11.0 dBm

 

 

 

20.0

– 40.0 GHz

+6.0 dBm

+3.0 dBm

 

 

 

 

 

 

 

 

68X75B

0.5 – 50.0 GHz

Standard

Standard

 

 

 

 

 

 

 

 

 

68X77B

0.01

– 50.0 GHz

Standard

Standard

 

 

 

 

 

 

 

 

68X85B

0.5 – 60.0 GHz

Standard

Standard

 

 

 

 

 

 

 

 

 

68X87B

0.01

– 60.0 GHz

Standard

Standard

 

 

 

 

 

 

 

 

68X95B

0.5 – 65.0 GHz

Standard

Not Available

 

 

 

 

 

 

 

 

 

68X97B

0.01

– 65.0 GHz

Standard

Not Available

 

 

 

 

 

 

 

 

682XXB/683XXB MM

3-7

Page 67
Image 67
Anritsu 683XXB, 682XXB manual Not Available 40.0 50.0 GHz DBm

682XXB, 683XXB specifications

The Anritsu 683XXB and 682XXB series are advanced vector network analyzers (VNAs) renowned for their precision and versatility in characterizing RF and microwave components. Designed for engineers and technicians involved in the development, manufacturing, and testing of high-frequency devices, these analyzers offer state-of-the-art technology that ensures optimal performance in various applications.

One of the hallmark features of the Anritsu 683XXB and 682XXB is their high dynamic range, which allows for accurate measurements of small reflection and transmission coefficients, essential for assessing the performance of complex RF structures. With frequency coverage extending from DC to 70 GHz, these analyzers cater to a broad spectrum of applications, making them suitable for industries such as telecommunications, aerospace, and automotive.

The user-friendly interface of the Anritsu VNAs is complemented by a high-resolution display, which facilitates easy navigation through measurement setups and results. The analyzers feature multiple measurement modes, including S-parameter measurements, time-domain analysis, and noise figure measurements, providing engineers with comprehensive tools for device characterization.

Both the 683XXB and 682XXB implement advanced calibration techniques, including automated calibration and error correction methods, to enhance measurement accuracy. These methods significantly reduce the uncertainties associated with test setups, enabling reliable performance evaluations of components like filters, amplifiers, and antennas.

Anritsu’s proprietary technologies, such as the VectorStar and ShockLine series integration, further empower the 683XXB and 682XXB models. These technologies enable high-throughput testing and improved measurement stability, addressing the needs of modern production environments that demand rapid turnaround times without sacrificing precision.

Additionally, the analyzers come equipped with various connectivity options, including USB, LAN, and GPIB, ensuring seamless integration into automated test systems. This adaptability enhances the analyzers' utility in both laboratory settings and field operations.

In conclusion, the Anritsu 683XXB and 682XXB series vector network analyzers represent the pinnacle of RF and microwave testing technology. With their unmatched precision, comprehensive measurement capabilities, and advanced calibration techniques, these instruments are indispensable tools for professionals striving to push the boundaries of high-frequency device performance and reliability.