682XXB/683XXB MM A-111

4-13 FM Calibration (68259B/68359B with Firmware Version 1.01 to 2.10)

Procedure Step Step Completion
1. FM Meter Calibration (calterm 123) ............................ ______________
2. FM Offset Balance Calibration (calterm 126) ....................... ______________
3. External FM Wide Sensitivity Calibration (calterm 124) .................. ______________
4. External FM Narrow Sensitivity Calibration (calterm 125) ................. ______________
5. Internal FM Wide Sensitivity Calibration (calterm 143) .................. ______________
6. Internal FM Narrow Sensitivity Calibration (calterm 144) ................. ______________
7. Store the Calibration Data ................................. ______________

4-13 FM Calibration (68259B with Firmware Version 2.18 and above;

68359B with Firmware Version 2.20 and above)

Procedure Step Step Completion
1. FM Meter Calibration (calterm 123) ............................ ______________
2. FM Variable Gain Linearity Calibration (calterm 148) ................... ______________
3. External Unlocked Wide FM Mode Sensitivity Calibration (calterm 124) ......... ______________
4. Internal Unlocked Wide FM Mode Sensitivity Calibration (calterm 143) .......... ______________
5. External Locked, Locked Low-Noise, and Unlocked Narrow FM Mode Sensitivity
Calibration(calterm 125) .................................. ______________
6. Internal Locked, Locked Low-Noise, and Unlocked Narrow FM Mode Sensitivity
Calibration(calterm 144) .................................. ______________
7. FM Rear Panel Input Gain Calibration (calterm 154) ................... ______________
8. Store the Calibration Data ................................. ______________
TEST MODEL
RECORD 68259B/68359B