A-200 682XXB/683XXB MM
TEST MODEL
RECORD 68277B/68377B
3-14 Pulse Modulation Tests: Rise Time, Fall Time, Overshoot, and Level (Continued)
Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 ms) Lower Limit Measured Value Upper Limit
SetF1 to 5.0 GHz
Setpulse width W1 to 5.0 µs
Recordthe displayed output level L1 (step 5.b) ....................
Recordthe displayed output level L1 (step 5.d) ....................
_______dBm
_______dBm
Calculateand record the pulse level error .............. –1.0 dB _______dB +1.0 dB
Pulse Level Accuracy (5 GHz, Pulse Width = 5.0 ms)
SetF1 to 5.0 GHz
Setpulse width W1 to 5.0 µs
Recordthe displayed output level L1 (step 5.b) ....................
Recordthe displayed output level L1 (step 5.d) ....................
_______dBm
_______dBm
Calculateand record the pulse level error .............. –0.5 dB _______dB +0.5 dB
3-15 Pulse Modulation Test: Video Feedthrough
Test Procedure Lower Limit Measured Value Upper Limit
SetF1 to 5.0 GHz
Setpulse width W1 to 5.0 s
SetPRI (period) to 0.01 ms
Measureand record the Video Feedthrough voltage spikes .... –10 mV _______ mV +10mV