9. On the Spectrum Analyzer:
a. Deselect NOISE LVL.
b. Set Frequency Span to 300 kHz.
c. Adjust the Marker for a 100 kHz offset.
d. Select NOISE LVL.
10. Measure the phase noise level 100 kHz offset
from the carrier frequency.Record the level on
the TestRecord.
11. Repeatsteps 1 through 10 for all frequencies
listed on the TestRecord.
682XXB/683XXB MM 3-25
PERFORMANCE SINGLE SIDEBAND
VERIFICATION PHASE NOISE TEST