A-136 682XXB/683XXB MM

3-9 Harmonic Test: RF Output Signals From 2.2 to 20 GHz (Continued)

Test Procedure (11 to 20 GHz) Measure Value Upper Limit
SetF1 to 12.4 GHz
Recordthe level of all harmonics of the 12.4 GHz carrier:
24.8GHz (2nd harmonic) .....................
37.2GHz (3rd harmonic) ......................
_______________dBc
_______________dBc
–60dBc*
–60dBc*
SetF1 to 16.0 GHz
Recordthe level of all harmonics of the 16.0 GHz carrier:
32.0GHz (2nd harmonic) ..................... _______________dBc –60dBc*
SetF1 to 20.0 GHz
Recordthe level of all harmonics of the 20.0 GHz carrier:
40.0GHz (2nd harmonic) ..................... _______________dBc –60dBc*
*–50 dBc if Option 15B (High Power) installed.
3-10 Single Sideband Phase Noise Test
Test Procedure Measured Value Upper Limit
SetF1 to 0.6 GHz
Recordthe phase noise levels at these offsets:
100Hz ...............................
1kHz ..............................
10kHz ..............................
100kHz ..............................
_______________dBc
_______________dBc
_______________dBc
_______________dBc
–84dBc
–97dBc
–95dBc
–112dBc
SetF1 to 2.0 GHz
Recordthe phase noise levels at these offsets:
100Hz ...............................
1kHz ..............................
10kHz ..............................
100kHz ..............................
_______________dBc
_______________dBc
_______________dBc
_______________dBc
–78dBc
–91dBc
–89dBc
–106dBc
TEST MODEL