0.5 - 2.2 GHz
Test
Procedure
The following procedure lets you measure the 0.5 to
2.2 GHz RF output harmonic levels to verify that
they meet specifications.
1. Set up the 682X5B/683X5B as follows:
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset the CW Menu is dis-
played.
b. Press Edit L1 toopen the current power level
parameter for editing.
c. Set L1 to the lesser of +10 dBm or the maxi-
mum leveled power level for the instrument
being tested (refer to Table3-2, page 3-6).
d. Press Edit F1 toopen the current frequency
parameter for editing.
e. Set F1 to the frequency indicated on the Test
Record.
2. Set up the Spectrum Analyzer as follows:
a. Span: 5 kHz/div
b. CF: Set to the 682X5B/683X5B frequency
value.
c. RBW: 1 kHz
d. Video Filter Wide: On
3. Adjust the Spectrum Analyzer Peaking control
for maximum signal level, then adjust the Refer-
ence Level Control to place the signal at the top
of the screen graticule.
4. Change the Spectrum Analyzer CF to each of the
harmonic frequencies listed on the TestRecord
and record the signal levels on the TestRecord.
Refer to Table3-3 (page 3-15) for the specified
harmonic signal level limits.
5. Repeat steps 1 through 4 for each of the 682X5B/
683X5B CW carrier and harmonic frequencies
listed on the TestRecord. Record the harmonic
signal levels on the TestRecord.
682XXB/683XXB MM 3-17
PERFORMANCE SPURIOUS SIGNALS TEST: RFOUTPUT SIGNALS
VERIFICATION £2 GHz (£2.2 GHzfor 68XX5B MODELS)