682XXB/683XXB MM A-45
3-11 Power Level Accuracy and Flatness Tests (Continued)

(Model 68247B/68347B with Option 15B High Power & without Option 2A Step Attenuator)

Power Level Accuracy Test Procedure
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power *
+13dBm
+12dBm
+11dBm
+10dBm
+9 dBm
+8 dBm
+7 dBm
+6 dBm
+5 dBm
+4 dBm
+3 dBm
+2 dBm
+1 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+17dBm
+16dBm
+15dBm
+14dBm
+13dBm
+12dBm
+11dBm
+10dBm
+9 dBm
+8 dBm
+7dBm
+6 dBm
+5 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
*Specification is 1.0dB. *Specification is 1.0dB.
Power Level Flatness Test Procedure (Step Sweep)
Set L1 to: Max Power Min Power Variation **
+13dBm _________dBm __________dBm _________dB
**Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 20 GHz).
Power Level Flatness Test Procedure (Analog Sweep) (Model 68347B only)
Set L1 to: Max Power Min Power Variation ***
+13dBm _________dBm __________dBm _________dB
***Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz)(typical, not a specification).
TEST MODEL
RECORD 68247B/68347B