6712FR Refrigerated Sampler Section 5 Extended Programming

 

Time Switched

 

Distribution

 

 

1

2 BOTTLES PER

SAMPLE EVENT (1-max)

 

 

•Enter the number of bottles that should receive samples at each sample event.

SWITCH BOTTLES ON:

NUMBER OF SAMPLES

2TIME

Select TIME.

3

SWITCH BOTTLES EVERY

__ HOURS, __ MINUTES

 

 

Enter the time between bottle switches.

FIRST SWITCH TIME

4AT: __:__

Enter the first bottle switch time.

If the program settings instruct the sampler to take a sample at the start time, the sampler also draws a sample at the switch time. When this occurs, the pacing will also be reset at the switch time.

Time switching is normally used with flow-paced programs that do not take a sample at start. These programs will place a sample in the current bottle if it is empty at the switch time, and then move to the next bottle/set. Pacing intervals are reset at every switch time.

When you use the time switching feature, the volume in each bottle or bottle set may vary. Sample volumes that would exceed the bottle capacity are not taken and a “PROBABLE OVERFILL” message is logged.

5.10.7Continuous Sampling Sample programs can be run indefinitely by selecting “YES” at the “RUN CONTINUOUSLY?” screen. Continuous sampling resets the distribution when the distribution sequence is com- plete. That is, when the last bottle/set is reached, the next bottle/set is the first bottle/set. All pacing modes except RANDOM are supported.

When running a program continuously, the bottles must be ser- viced at regular intervals to prevent overfilling the bottles. The sampler assumes that the next bottle/set is empty and ready to receive samples. If you are using this feature with two-part pro- gramming, keep in mind that the parts will most likely reset at different intervals.