Follow the instructions in Opening the Controller Case in Section

6712FR Refrigerated Sampler

Section 8 Maintenance

Figure 8-3 Opening the Controller CAUTION

Do not attempt to reactivate the used desiccant. Reactivation may cause desiccant to leak out of the bag, causing mechani- cal damage inside the controller.

Follow the instructions in Opening the Controller Case in Section

8.6to open the controller. Remove the bag from the cardboard box (part #69-9003-518) at the front of the case and replace it with a fresh bag of desiccant (part #099-0002-00).

Distributor

Connectors

Desiccant Box

Figure 8-3 Opening the Controller

8-13