Teledyne manual 1 6712FR Sampler Features Continued, 6712FR Refrigerated Sampler Introduction

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 22
Image 22
Extended Programming Features

6712FR Refrigerated Sampler

Section 1 Introduction

Table 1-1 6712FR Sampler Features (Continued)

Option for a Continuous Run- ning Program

Continuous Sampling: When sample bottles are regularly replaced, the sample distribution can restart with the first bottle set after the last bottle set is filled, without interrupting the running program.

Extended Programming Features

Pacing

Uniform Time Pacing: Sampling at regular time intervals.

 

 

Nonuniform Time Pacing: Sampling at irregular time intervals.

 

Random Interval Pacing: Sampling at unique random time intervals generated

 

by the controller.

 

Flow Pacing: Sampling at regular flow-volume intervals.

 

Event Pacing: Sampling each time a user-definable event occurs.

 

 

Flow Proportional

Allows for sample sizes to be based on flow. (This option is available only with

Sample Volumes

uniform time pacing.)

 

 

Distribution Methods

Composite: Samples deposited in a single large bottle.

 

 

Sequential: Only one sample placed in each bottle.

 

Samples Per Bottle: Multiple samples placed in each bottle

 

Bottles Per Sample: One sample deposited in multiple bottles

 

Multiple Bottle Compositing: A combination of samples per bottle and bottles

 

per sample distribution methods.

 

Time Switched Bottles or Bottle sets: Control the sample distribution using

 

clock times.

 

 

Programmable Sampler Enable

A 6712FR sampler can be programmed to enable or disable a running sampling

 

program when readings received from a connected rain gauge, module, or

 

SDI-12 Sonde meet certain conditions.

 

 

Pauses and Resumes

Create intermittent sampling schedules.

 

 

Two-Part Programming

Two-part programming lets you set up a sampling program that divides the bot-

 

tles into two groups, filling each group according to separate pacing, distribution,

 

sampler enable, and pause and resume settings. This is ideal for storm-water

 

run-off sampling.

 

 

Auto Suction Head or Fixed Suc-

The suction head, is the vertical distance from the flow stream to the liquid detec-

tion Head

tor. Extended programming has two settings.

 

Auto-Suction Head: The head is automatically determined.

 

Fixed Suction Head: A user-definable measurement for the head.

 

 

Suction Line Rinses

Program setting for the number of times (0 to 3) that the 6712 rinses the suction

 

line before drawing a sample.

 

 

Sampling Retries

Program setting for the number of times (0 to 3) that the 6712 attempts to sample

 

if it fails to deliver the entire sample volume.

 

 

Three Flexible

Run Immediately: Starts the sampling program immediately.

Start Time Settings

Delayed Start: Starts the sampling program after a user-definable delay of 1 to

 

 

999 minutes.

 

Clock Time: Starts the sampling program at a user-definable time on one or

 

more days of the week.

 

 

Option for a Continuous Run-

Continuous Sampling: When sample bottles are regularly replaced, the sample

ning Program

distribution can restart with the first bottle set after the last bottle set is filled,

 

without interrupting the running program.

 

 

1-6

Page 22
Image 22
Teledyne 1 6712FR Sampler Features Continued, 6712FR Refrigerated Sampler Introduction, Extended Programming Features