Teledyne manual 6712FR Refrigerated Sampler Safety

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 5
Image 5
6712FR Refrigerated Sampler

6712FR Refrigerated Sampler

Safety

The 6712FR Refrigerated Sampler is a “definite purpose” device, intended for use only with compatible Isco equipment. Do not use this product with any other manufacturers’ equipment, or for any other purpose. Use for any purpose not described in this manual could cause personal injury or property damage.

Electrical Requirements The refrigerator is available in both 120 VNever operate the refrigerator in an explosive atmo- sphere. and 230 VManual backgroundManual background config- urations. The required operating voltage for the refrigerator is listed on the Identification and Serial Number label, placed on the inside of the sample compartment door.

Refrigerators configured for 120 VManual backgroundManual background operation are equipped with a North American NEMA 5-15P plug and is intended for use only with 120 VManual backgroundManual background, 60 Hz. The power source should be rated for 20 ampere service.

Refrigerators configured for 230 VManual backgroundManual background operation are equipped with a Continental European CEE 7/7 plug and is intended for use only with 230 VManual backgroundManual background, 50/60 Hz. The power source should be rated for 16 ampere service.

Refrigerators in either configuration provide 12.5 VManual background at 5 amperes for the sampler controller. This output is from the two-pin military-type connector on the cord on top of the refriger- ator. This output is intended for 6700 Series Sampler controllers only.

The refrigerator must be installed near a suitable power outlet. Never use an extension cord.

The power outlet must be visible and easily accessible. Unplugging the refrigerator is the only means of discon- necting power.

To minimize the risk of electrical shock, the refrigera- tor must be connected to an outlet with an electrical ground contact.

The power source must be a dedicated circuit. The line must not power any other devices.

Never operate the refrigerator with the lower front or rear panels removed.

Never operate the refrigerator in an explosive atmo- sphere.

Do not locate the refrigerator where the lower compart- ment could become submerged.

Do not lift or carry the refrigerator. Use an appliance carrying device.

Fuses must be replaced with the required size, current, voltage, and blow-time specifications. Refer to the Replacement Parts List for the correct part number.

iii