Table 7-4 6712 Remote Phone Commands (Continued)

6712FR Refrigerated Sampler

Section 7 Remote Operation

Table 7-4 6712 Remote Phone Commands (Continued)

Phone

Name

Description

Command

 

 

5

#

Disable

This command is similar in functionality to a pin F low signal (a signal typically

 

 

 

used to enable or disable the sampler from a flow meter). At run time, the sam-

 

 

 

pler starts with the phone enable in the enabled state. PHONE DISABLE is

 

 

 

valid while a program is running. If successful, a reply of REMOTE ENABLE IS

 

 

 

FALSE will be returned.

 

 

 

A Remote Disable is the highest priority disable. If the sample is disabled with

 

 

 

this command, no condition other than command 6#, ENABLE, will allow the

 

 

 

program to become enabled again.

 

 

 

 

6

#

Enable

This command is used to re-enable a phone-disabled sampler. PHONE

 

 

 

ENABLE is valid while a program is running. If successful, a reply of REMOTE

 

 

 

ENABLE IS TRUE will be returned.

 

 

 

 

7

#

Grab Sample

This command causes the sampler currently running a program to take a sam-

 

 

 

ple. The sample will count as one of the program’s samples and is placed in

 

 

 

the current bottle (Part A’s current bottle when using two-part programming).

 

 

 

TAKE_SAMPLE is valid while the program is running, after the start time, but

 

 

 

not while in the MANUAL PAUSE screens. If successful, a response similar to

 

 

 

PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a

 

 

 

response similar to NO SAMPLE, PROGRAM WAITING START will be

 

 

 

returned.

 

 

 

 