2.2.1 Installing Racks

6712FR Refrigerated Sampler

Section 2 Installation/Preparation

2.2Installing Bottle Kits Teledyne Isco ships the refrigerator from the factory with the sample bottles in place. When using the sampler for the first time, you will only need to remove the bottle lids. When installing cleaned bottles or a new kit, keep these guidelines in mind:

Do not mix glass and plastic bottles together.

Install all bottles to insure that they remain in place.

Align the racks correctly. If the racks are misaligned, the sampler may miss the bottle mouth, or the sampler will deposit samples in the “wrong” bottle.

 

Each time you change from one size bottle to another, you must

 

adjust the length of the distributor arm, install a discharge tube

 

of the correct length, and possibly install a different distributor

 

shaft extension.

 

Each time you install a bottle kit, check the program settings for

 

the number of bottles and bottle volumes. Also rotate the dis-

 

tributor arm to each bottle position to ensure the arm is correctly

 

positioned over each bottle. Use the controller to move the arm.

2.2.1 Installing Racks

Install the bottle rack with bottles by sliding it into the cooling

 

compartment until the rack drops over the two semicircular

 

ramps which hold it in place (Figure 2-2).

 

The five posts and two ramps which position the rack under the

 

distributor arm are adjusted at the factory. When you change

 

bottle kits, you may need to readjust the stops and ramps and

 

install a different distributor arm extension. Figures 2-3and 2-4

 

show you how to install the bottle kits.

 

Check the kit’s alignment by rotating the distributor arm. Do

 

not rotate the distributor manually. Moving the arm man-

 

ually damages the distributor drive. If you see any mis-

 

alignment, adjust the posts and ramps until the discharge tube

 

at the end of the arm stops over each bottle.