6712FR Refrigerated Sampler

Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)

11

12

13

14

15

16

17

18

19

20

21

22

PROGRAM MODULE?

YES NO

MODE OF OPERATION

FLOWMETER

LEVEL ONLY

WEIR FLUME

DATA POINTS

EQUATION MANNING

FLOW-INSERT

WEIR/ORIFICE TYPE

V-NOTCH

ROUND

FLOW INSERT SIZE:

6" 8" 10" 12"

NEW MODULE SETUP-- DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

CURRENT LEVEL IS

___.___ ft.

ADJUST LEVEL TO

___.___ ft.

DATA STORAGE

INTERVAL IN MINUTES

1

2

5

10

15

30

NUMBER OF BOTTLES:

1 2 4 8 12 24

BOTTLE VOLUME IS 1000 ml (300-30000)

SUCTION LINE LENGTH

IS 5 ft (3-99)

AUTO SUCTION HEAD

ENTER HEAD

Select YES

Select FLOWMETER

Select FLOW-INSERT.

Select ROUND.

Select 12.

Select DONE.

Note: Partition memory status message will be displayed briefly.

Press ENTER.

Select 5. (If the data storage interval is different from the current program, you will be informed that the storage interval has changed, and that the stored data will be lost. Download stored data (if desired) and select DONE.

Select 24.

Enter 1000.

Enter 5. The 6712FR calculates new pump tables.

Select AUTO SUCTION HEAD.

5-7

Page 95
Image 95
Teledyne 6712FR manual Flow-Insert