When you are done validating, select “DONE” to return to the

6712FR Refrigerated Sampler

Section 6 SDI-12 Sondes

Use the arrow keys to select the desired parameter to validate. Any parameter can be validated. You will be prompted to place the probe in your validation standard. When stable, press ↵. The reading will be displayed at the bottom of the screen.

When you are done validating, select “DONE” to return to the

Hardware options screens.

6.7Disconnecting Sondes When you have disconnected all sondes, and have interrogated the sampler for all its data, you can indicate that there is no SDI-12 sonde.

From the main menu, select Other Functions, Hardware. Advance through the screens until you see one that displays “SDI-12 DATA.” Select this screen to display the following:

NO SDI-12 SONDE

SETUP KNOWN SONDES

SCAN SDI-12 SONDES

Select “NO SDI-12 SONDE.” You will advised that you have a new hardware setup:

NEW HARDWARE SETUP--

DOWNLOAD DATA NOW

OR LOSE ALL DATA!

DONE

If you have collected all the data, select “DONE.” A partition memory message will be displayed briefly, and you will be returned to the Hardware options screens.

If you do not have a sonde attached, you may want to turn off automatic scanning for sondes. This will eliminate the time taken to scan for sondes whenever the sampler is turned on.

6-7