Teledyne 6712FR manual 2 Storm Water Runoff Sampling Continued, 5-11, Yes No, Set Point, 0.25 ft

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 99
Image 99
5-11

6712FR Refrigerated Sampler

Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)

61

62

63

RESET RAIN HISTORY?

YES NO

“LEVEL” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

“LEVEL” SET POINT

0.25 ft

(0.100-30.000)

Select YES.

Select SET POINT.

Enter 0.25.

64

65

66

67

68

69

70

ONCE ENABLED,

STAY ENABLED?

YES NO

SAMPLE AT ENABLE?

YES NO

START OF SAMPLING

AFTER A

15 MINUTE DELAY

(1-999)

PAUSE RESUME

1.HH:MM DD HH:MM DD

2.HH:MM DD HH:MM DD CLEAR DONE

NO DELAY TO START

DELAYED START

CLOCK TIME

WAIT FOR PHONE CALL

START “STORM”

AFTER A 72 HOUR

“DRY” PERIOD?

YES NO

START “STORM”

AFTER A

1 MINUTE DELAY

(1-999)

Select YES.

Select NO.

Enter 20.

Select DONE.

Screen will say “End of Part B.”

Select DELAYED START.

Select NO. (Option for samplers using a 0.01 inch tip rain gauge. See Dry Period Option, described on page 5-26.)

Enter 1.

71

PROGRAMMING COMPLETE

RUN THIS PROGRAM

NOW?

YES NO

Select NO.

The above example presented several screens that are not available in standard programming. Descriptions of features available in extended programming follow.