Teledyne 6712FR Refrigerated Sampler, Section, Standard Programming, 4.12.1, Calibration Tips

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 72
Image 72
4.12.1

6712FR Refrigerated Sampler

 

Section 4

Standard Programming

 

 

 

 

4.12.1

Calibration Tips

• For the best results, calibrate the sampler after it’s

installed on site.

• Use a graduated cylinder for volume measurement.

• The sampler clears the previous calibration setting when you reinitialize the sampler.

• You cannot calibrate while the sampler is running a program.

• When the sampler delivers the sample, it runs through a complete sampling cycle, using the current settings for volume line rinses and retries. If it is a two part program, the sampler uses the volume setting for part A.

Calibrating Sample Volumes

 

RUN

1

PROGRAM

VIEW REPORT

 

OTHER FUNCTIONS

 

 

•Select OTHER FUNCTIONS

 

MAINTENANCE

2

MANUAL FUNCTIONS

MODULE

 

HARDWARE

 

 

•Select MANUAL FUNCTIONS

 

GRAB SAMPLE

3

CALIBRATE VOLUME

OPERATE PUMP

 

MOVE DISTRIBUTOR

 

 

•Select CALIBRATE VOLUME

4

SAMPLE VOLUME:

___ ml (100-9990)

 

 

•Enter the desired volume

CALIBRATE VOLUME

5PRESS WHEN READY!

 

VOLUME DELIVERED:

9

____ml

 

 

 

•Measure the volume delivered and enter the amount here.

 

200 ml

10

ARE YOU SURE?

YES NO

 

 

This screen appears when the amount you entered in step 9 was more than twice, or less than half, the programmed volume.

Select YES when the volume delivered matches the volume displayed on this screen.

Select NO to enter the volume delivered.

To calibrate:

1.Disconnect the pump tube from the bulkhead fitting.

2.Place the end of the tube over a collection container.

3.Follow the steps in Calibrating Sample Volumes in the margin.

6

PURGING

SUCTION LINE

Reconnect the pump tube to the bulkhead fitting.

7

PUMPING 200 ml

8

PURGING

SUCTION LINE

4-16

Page 72
Image 72
Teledyne manual 6712FR Refrigerated Sampler, Section, Standard Programming, 4.12.1, Calibration Tips