Teledyne manual 6712FR Refrigerated Sampler Standard Programming, Collecting Reports

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 74
Image 74
4.15 Reports

6712FR Refrigerated Sampler

Section 4 Standard Programming

4.15 Reports

The 6712 records a variety of data while running a sampling program. It uses the data to produce four reports:

The Program Settings report, listing the program settings.

• The Sampling Results report, listing the program settings, time of samples and other program events.

Interrogator

The Combined Results report, combining the sample event times with rainfall or module data.

The Summary report, listing daily summaries of data collected. The Summary report may be rainfall and/or module data.

The sampler stores the data for each report in memory where they remain until you select RUN. Selecting RUN clears the memory so that it can store the data from the next program.

4.15.1 Collecting Reports

There are a number of ways to collect the reports:

 

View reports (program settings, sampling results, and abbre-

 

viated summary reports) on the sampler’s display.

 

Collect the reports (except for summary reports) with an Isco 581

 

RTD and use a computer running flowlink or samplink to

 

transfer the reports from the RTD to a file on the computer.

 

Collect the reports (except for summary reports) with an IBM PC

 

or compatible computer running flowlink.

581 RTD

Collect the reports with and IBM PC or compatible computer

running samplink.

 

4.15.2 Viewing the Data

Because the display area is small, the data on the reports will be displayed a little differently from the data that is transferred to and viewed on your computer.

• For the Sampling Report, the sampler displays each program event, one at a time.

• For the Module and Rainfall Data reports, it displays daily summaries, instead of the full reports available with the rtd, Flowlink, or Samplink.

If you select Sampling Report or Rainfall (step 3 in Viewing Data, shown in the margin), the sampler begins displaying the report data. The sampler advances automatically through the report items, displaying each item briefly. While the sampler advances automatically through the displays:

• Stop the automatic displays by pressing Stop once. Then, use the arrow keys to move manually through the report.

4-18

Page 74
Image 74
Teledyne manual 6712FR Refrigerated Sampler Standard Programming, Collecting Reports, Viewing the Data