Delayed Start Screen

6712FR Refrigerated Sampler Section 4 Standard Programming

Delayed Start Screen

NO DELAY TO START

DELAYED START

1CLOCK TIME

WAIT FOR PHONE CALL

• Select DELAYED START.

FIRST SAMPLE

AFTER A

2___MINUTE DELAY

(1-999)

Enter the number of minutes to delay the countdown to the first sample.

START FLOW COUNT

AFTER A

3__MINUTE DELAY

(1-999)

The sampler uses this screen for flow paced programs.

Clock Start Time Menu

NO DELAY TO START

DELAYED START

1CLOCK TIME

WAIT FOR PHONE CALL

Select CLOCK TIME.

FIRST SAMPLE AT

206:00

Enter the time of day.

SELECT DAYS:

SU MO TU WE TH FR SA

3

DONE

Press an arrow key until the cursor is on

the correct day. Press.

Repeat until each day you want is blinking.

Press an arrow until DONE blinks. Press .

4.6.2 Sampler

Isco flow meters and flow loggers have a programmable

Enable/Disable

sampler-enable feature that lets them enable (start) or disable

 

(stop) a program according to certain monitored conditions. For

 

example these conditions can be level, flow rate, pH, temper-

 

ature, percent, rainfall, I/O; or a combination of two conditions.

 

The Model 1640 Liquid Level Actuator can also be used, as well

 

as any other instrument that can ground the input to pin F.

 

Grounding the input disables the sampler.

 

The sampler disregards disable signals during the delay to start

 

time. However, when the sampler detects a disable signal at the

 

start time, it suspends the program until it is enabled. The

 

diagram in Figure 4-2shows the sampler’s responses.