4.15.4

6712FR Refrigerated Sampler

 

Section 4

Standard Programming

 

 

 

 

4.15.4

Sampling Results

As the sampler runs a program, it records the program’s events.

 

Report

Program events include such items as sample events, program

 

 

enables and disables, power losses, and so on. Table 4-3lists each

 

 

event with a short description. You can view the report by fol-

 

 

lowing the steps in Viewing the Data on page 4-18or print the log

 

 

as the Sampling Results report.

 

 

In the text file version of the report, sample-event entries include

 

 

the time of the event as well as the sample and bottle numbers.

 

 

The sample column contains the sample number deposited into

 

 

the bottle. This information reflects the distribution selected for

 

 

the program.

 

 

The letters in the Source columns are codes for the cause of the

 

 

event. The letters in the Error column are codes for the causes of

 

 

missed samples. The source and error codes appear with expla-

 

 

nations at the end of the report. A list of codes appears in Table

 

 

4-3.

 

 

The last column in the Sampling Results report records the

 

 

pump counts for the sample event. Large variations in pump

 

 

counts from event to event indicate fluctuating heads, a relocated

 

 

suction line, or a temporarily clogged strainer.

4.15.5

Combined Results

At the time of each sample event, the sampler records the

 

Report

readings from each connected module, rain gauge, or SDI-12

 

 

sonde. The Combined Results Report presents the event time,

 

 

sample and bottle number, and the respective readings from the

 

 

device.

 

 

For flow modules, the readings are level and flow rate. For the

 

 

pH module, the readings are temperature and pH. For the rain

 

 

gauge and SDI-12 Sonde, the readings vary according to the

 

 

selections enabled during the hardware setup in extended pro-

 

 

gramming. When recording four or more sonde parameters, the

 

 

Combined Results report will exceed the standard 40 character

 

 

width. To read the additional columns, the data must be collected

 

 

with Flowlink 4 for Windows software.

4.15.6

Summary Report

This report summarizes the rain gauge or module data. The

 

 

types of readings are the same as those available to the Com-

 

 

bined reports. The sampler compiles the data to present a daily

 

 

total, the day’s average, maximum, and minimum reading, and a

 

 

listing of hourly averages. The sampler then graphs the day’s

 

 

readings.

4-20

Page 76