Section 7 Remote Operation

 

 

6712FR Refrigerated Sampler

 

 

Section 7 Remote Operation

 

 

 

 

 

Table 7-1 6712 Remote Menu Commands

 

 

 

Menu Command

Description

 

0 or START

This command will immediately start any program set to WAIT FOR PHONE CALL and cur-

 

rently waiting for the remote start command (see Section 4.6).

 

 

1 or RUN1

This command will load Program 1 as the sampler’s current Extended program, reset partition if

 

the storage interval has been changed, and run the program. RUN1 is valid when the sampler

 

is in standby, or when it is waiting for a start command. RUN1 will also run the current program

 

while in Standard programming.

 

 

2 or RUN2

These commands will load the selected program number as the sampler’s current Extended

3 or RUN3

program, reset partition data if the storage interval has changed, and run the program. They are

valid when the sampler is in standby, or when it is waiting for a start command. The commands

 

4 or RUN4

are available for Extended programming only.

 

 

5 or DISABLE

This command is similar in functionality to a pin F low signal (a signal typically used to enable or

 

disable the sampler from a flow meter). At run time, the sampler starts with the sampler in the

 

enabled state. DISABLE is valid while a program is running. If successful, a display of REMOTE

 

ENABLE IS FALSE will be returned.

 

A Remote Disable is the highest priority disable. If the sampler is disabled with this command,

 

no condition other than menu command 6, ENABLE, will allow the program to become enabled

 

again.

 

 

 

6 or ENABLE

This command is used to re-enable a disabled sampler. ENABLE is valid while a program is

 

running. If successful, a display of REMOTE ENABLE IS TRUE will be returned.

 

 

7 or TAKE_SAMPLE

This command causes the sampler currently running a program to take a sample. The sample

 

will count as one of the program’s samples and is placed in the current bottle (Part A’s current

 

bottle when using two-part programming). TAKE_SAMPLE is valid while the program is run-

 

ning, after the start time, but not while in the MANUAL PAUSE screens. If successful, a

 

response similar to PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a

 

response similar to NO SAMPLE, PROGRAM WAITING START will be returned.

 

 

ST or STATUS

This command causes the sampler to return its status information and current parameter data.

 

 

SD or

This command causes the sampler to return the information currently displayed on the sam-

SCREEN_DUMP

pler’s screen. The remote screen follows the sampler’s display as the information is updated.

 

This mode can be used to observe from a remote location while the sampler is programmed

 

on-site.

 

 

 

P or PAUSE

This command causes a sampler currently running a program to enter the MANUAL PAUSE

 

menu. While you are connected in this mode, the sampler keypad becomes disabled, and your

 

computer’s keypad will be redirected to simulate the sampler keypad. The following keys will be

 

activated:

 

 

<Enter>, decimal point, and number keys.

 

<Esc>, S, or s = STOP.

 

<Backspace>, L or l, and U or u = Left/Up arrows.

 

R or r, and D or d = Right/Down arrows.

 

 

 

Sampling reports

You can retrieve the sampling reports at any time using the same

 

 

serial connection. Type “Q” to exit the menu. Then at the >

 

 

prompt, type “REPORT” and press Enter. The sampler controller

 

 

will send reports as configured.