Table 5-3 Event Paced Sampling (Continued)

6712FR Refrigerated Sampler

Section 5 Extended Programming

Table 5-3 Event Paced Sampling (Continued)

5

6

7

8

9

10

11

12

13

14

15

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

EVENT02:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 DONE

“LEVEL” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

CONDITION IS TRUE

WHEN “LEVEL”

RISES FALLS

“LEVEL” RISES

1.000 ft

__HOURS, __MINUTES

EVENT03:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 DONE

“pH5” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

“pHx” RANGE: LOWER: 4.00 UPPER: 10.00 ( 0.00 - 14.00)

ENABLED WHEN:

INSIDE RANGE

OUTSIDE RANGE

EVENT04:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 DONE

PACING:

EVENT PACED

3 EVENT CONDITIONS

Select ABOVE SET POINT.

Select LEVEL again.

Choose RATE OF CHANGE.

Choose RISES.

This will enable the event condition if a rise occurs within the time frame and above the amount specified in screen #9.

Enter 1 ft and press .

Then enter 2 hours.

Select pH.

Select RANGE.

Enter 4 for the lower range and 10 for the upper range.

Choose INSIDE RANGE.

You will not be entering a fourth event condition, so use the arrow keys to scroll down, and then select DONE.

Proceed with the rest of the programming steps.

5-14

Page 102
Image 102