Teledyne 6712FR manual 2 Storm Water Runoff Sampling Continued, Select RAIN

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 97
Image 97
Select RAIN.

6712FR Refrigerated Sampler

Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)

35

36

37

38

39

40

41

42

43

44

45

46

47

ENABLE:

RAIN LEVEL FLOW

NONE

ENABLE: RAIN

AND OR DONE

ENABLE: RAIN AND

LEVEL FLOW NONE

“RAIN” SET POINT: 0.15 INCHES PER __HOURS, __MINUTES

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

RESET RAIN HISTORY?

YES NO

“LEVEL” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

“LEVEL” SET POINT

0.25 ft

(0.001-30.000)

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

ONCE ENABLED,

STAY ENABLED?

YES NO

SAMPLE AT ENABLE?

YES NO

PAUSE RESUME

1.HH:MM DD HH:MM DD

2.HH:MM DD HH:MM DD CLEAR DONE

UNIFORM TIME PACED

FLOW PACED

EVENT PACED

NONUNIFORM TIME

Select RAIN.

Note: All valid enable options will appear here, so your screen may differ.

Select AND.

Select LEVEL.

Enter 0.15 for inches, then enter 30 minutes.

Select ABOVE SET POINT.

Select YES.

Select SET POINT.

Enter 0.25.

Select ABOVE SET POINT.

Select YES.

Select YES.

Select DONE.

Screen will say “End of Part A” then “Beginning Part B.”

Select FLOW PACED.

5-9