Table 4-4 Programming Example With 730 Module Installed

6712FR Refrigerated Sampler

Section 4 Standard Programming

4-29 CAUTION

Most Isco accessories and interfacing equipment are not certi- fied for us in hazardous locations as defined by the National Electrical Code. Check applicable guidelines, codes, and regu- lations of federal, state, city, and county agencies. Refer to Appendix C in the back of this manual for more safety informa- tion.

Table 4-4 Programming Example With 730 Module Installed

1

2

MODULE INSERTED---

DOWNLOAD DATA NOW

OR LOSE ALL DATA!

DONE

6712 SAMPLER

STANDARD PROGRAMMING

For HELP at any

screen press ? key.

With the sampler off, insert the module in the bay on the controller. Turn the sampler on by pressing the On/Off key. Press .

Download stored data, if you want to keep it. Pressing DONE erases it and advances to the next screen.

This screen will now appear for 8 seconds.

3

RUN

The option PROGRAM will be blinking. Press .

PROGRAM

 

VIEW REPORT

 

OTHER FUNCTIONS

 

 

 

4

5

6

7

8

9

SITE DESCRIPTION:

“FACTORY051“

CHANGE?

YES NO

SELECT UNITS FOR

FLOW RATE:

cfs gps gpm Mgd lps m3s m3h m3d

SELECT UNITS FOR

FLOW VOLUME:

cf gal Mgal

m3 lit

PROGRAM MODULE?

YES NO

MODE OF OPERATION:

FLOWMETER

LEVEL ONLY

WEIR FLUME

DATA POINTS

EQUATION MANNING

FLOW-INSERT

The option NO will be blinking. Press .

For the purposes of this example, it isn’t necessary to change the description.

Select the desired units for the flow rate. For this example, choose cfs. Press .

Select the desired units for the flow volume. For this example, choose Mgal. Press .

Select YES to program the module.

If you choose Flowmeter, proceed to screen #9 below.

If you choose LEVEL ONLY, proceed to screen #12.

When you select Flowmeter, you will have several options for performing the flow conversion.

For this example, choose Weir. Press .