Table 1-1 6712FR Sampler Features

6712FR Refrigerated Sampler

Section 1 Introduction

nected to the optional digital cellular modem), or from a 581 Rapid Transfer Device (RTD). The RTD is a quick and simple way to transfer the data from the field to your computer. Reports are easily collected with Teledyne Isco’s Samplink software.

1.5Pump Requirements The pump also requires a pump tube made specifically for the 6712 and 6700 Series samplers. These pump tubes are easily rec- ognized by their blue alignment collars. The 6712 pump tubing is the same as that for Teledyne Isco’s 6700 Series samplers, but different from Teledyne Isco’s earlier model samplers, such as the 3700 Series. Other types of pump tubing will not work in the 6712FR. Refer to Replacing the Pump Tube on page 8-9.

General Features CAUTION

This sampler has a high performance pump. As with all such pumps, it relies upon liquid to cool working components. If the sampler is programmed to pump in the absence of liquid in excess of 5 minutes, excessive heat buildup may damage the paddles, rollers, and housing. Ensure that the liquid inlet is completely immersed.

Table 1-1 6712FR Sampler Features

General Features

Top Cover

• Protects pump, power source, and controller.

 

 

• Lockable latches.

 

 

Controller Only

• Contains a rechargeable desiccant to prevent moisture damage to the

 

(does not include refrigerator)

electronics, pump, and distributor systems.

 

• Control panel sloped 15 degrees for easy reading.

 

• Keys labeled with large, vivid icons.

 

80-character display (4 lines by 20 columns).

 

• Display has selectable backlight:

 

Always on or always off.

 

Timed, switching off when keypad is inactive for 60 seconds.

 

• Memory for program and data storage.

 

• Flash memory for easy software upgrades.

 

• NEMA 4X and 6 (IP67) ratings.

 

 

Refrigerator

• The 6712FR requires 120 volts AC, 60 Hz, or optional 230 volts Ac, 50 Hz. A

 

 

built-in 12 volt DC power converter powers the controller.

 

• The power supply and solid state thermostat are sealed inside the refrigerator’s

 

base. However, electrical connections for the fan and compressor are not

 

sealed.

 

• A forced air condensing coil and front ventilation let you place the unit close to a

 

wall or in a corner.

 

• The oversized wrap-around evaporator plate cools the sampling compartment

 

quickly and efficiently. Heaters on the plate let the sampler continue to operate

 

in cold temperatures. The evaporator plate is self-defrosting.

 

Food-grade ABS plastic interior will not support bacterial growth or leach

 

plasticizers into the sample.

 

 

1-3

Page 19