Power Consumptiona

6712FR Refrigerated Sampler

Section 1 Introduction

Table 1-3 Technical Specifications for the 6712FR Refrigerated Sampler (Continued)

 

 

Power Consumptiona

 

6712 Controller in Standby

 

 

10 mA

 

 

 

 

Average Current of Accessories

 

 

 

 

 

 

Controller Display Backlight

 

228 mA

 

 

 

 

701 pH Moduleb

 

 

11 mA

710 Ultrasonic Moduleb

 

11 mA

720 Submerged Probe Moduleb

 

11 mA

730 Bubbler Moduleb

 

 

— at 2 ft of liquid level

 

24 mA

— at 10 ft of liquid level

 

31 mA

 

 

 

750 Area Velocity Moduleb

 

12 mA

780 4-20 mA Moduleb

 

10 mA

YSI 600 with pH, DO, Conductivity, and Temperatureb

11 mA

Programmable Analog 4-20 mA Output Optionc

 

124 mA

CDMA cellular telephone modem

 

 

— standby current

 

 

35 mA

— current while communicating

 

150 mA

 

 

 

GSM cellular telephone modem

 

 

— standby current

 

 

62 mA

— current while communicating

 

230 mA

 

a. Current ratings when input power is 12.5 VDC and ambient temperature is 77 °F (25 °C)

b. Average current when programmed to take readings at 15 minute intervals.

 

c. Rating for three outputs.

 

 

 

 

 

Sampler Only

 

Approximately 30 mA based on 200 ml sample every hour, 10 ft (3 m) suction

Operating Current:

 

line, and 5 ft (1.5 m) suction head.

 

 

 

 

Isco External Battery Capacity:

Number of Complete Standard Sampling Routines

 

 

Nickel-Cadmium Battery

Lead-Acid Battery

Sampler Only:

 

6

9

Sampler with pH Module:

 

6

9

Sampler with Ultrasonic Module:

6

9

Sampler with Submerged Probe Module:

6

9

Sampler with Bubbler Module (1-2 ft head):

3

5

Sampler with Area Velocity Module:

4

6

Sampler with 4-20 mA Module:

6

9

A Standard Sampling Routine is a 200 ml sample taken every hour in 24 bottle mode, with a 10 foot (3 m) vinyl suc- tion line at 5 feet (1.5 m) of head. The module is set to take a reading every 15 minutes.