4.8Interrupting a Running Program

6712FR Refrigerated Sampler Section 4 Standard Programming

BOTTLE 2

IN 00:12:00

***** gpm

5.73 ft

00001457600 gal

4.8Interrupting a Running Program

4.8.1Stop Program

4.8.2Resume Program

An asterisk (*) appears next to the reading if the module was unable to take a reading. If an asterisk appears, the reading dis- played is the last available reading. Flow rates will be reported as zero flow if the error persists for more than five minutes (for a level or velocity reading that flow is calculated from).

If an entire reading is filled with asterisks, the value exceeds the number of characters that can be displayed. The sampler stores the actual reading in memory, so it may be collected later.

For those programs that have delayed or scheduled start times, the readings will be displayed while waiting for the start time. Keep in mind that any totalizer values will be reset at the start time. Module and sonde readings are also displayed after a program is complete, until a key is pressed.

It is not necessary to run a program to view module or sonde readings. Real time data can be viewed on the display by selecting “VIEW REPORT” from the main menu.

You can interrupt a sampling program by pressing the Stop key while the sampler is waiting for the next sample event. Pressing Stop places the sampler into Manual Paused operation and records a manual pause in the sample event log.

In this state, the sampler allows you to access several functions. You can modify the program, reset alarms, take a grab sample, etc., without adversely affecting the running program. Modifying the program is limited to adjusting the pacing intervals, enable conditions and sample volume. The basic program structure, such as the type of pacing, can not be changed.

While in the manual paused state, the sampling program con- tinues to operate as normal, with the exception of taking samples. If a sample was to be taken, it is skipped. The sampler records “sample skipped” in the sample event log and continues to operate as normal.

The manual paused state displays a scrolling menu with several options (see example below). Use the Arrow keys to scroll through the manual paused options and the ↵ key to make a selection. Return to the running program by selecting resume program or pressing the Stop key. If you do not make a selection, the sampler automatically resumes the original sampling program five minutes after the last key was pressed.

STOP PROGRAM

Manual Paused Screen

RESUME PROGRAM

VIEW DATA

 

GRAB SAMPLE

 

 

 

This option terminates the running program and records program stopped in the sample event log. The program cannot be resumed.

Selecting this option will cause the sampler to exit the manual paused state and return to normal program operation. A manual resume event will be logged.