Teledyne 4 Sampler Enable, 6712FR Refrigerated Sampler, Section, Extended Programming, 5.12.1

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 112
Image 112
Extended Programming

6712FR Refrigerated Sampler

 

Section 5

Extended Programming

 

 

 

 

5.12.1

Sampler Enable

This example uses a sampler with an attached 730 Bubbler Flow

 

 

Module, Rain Gauge and an SDI-12 sonde at address 5 with four param-

 

 

eters.

1

2

3

4

5

6

7

8

9

10

11

ENABLE:

NONE PROGRAMMED

ENABLE:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 NONE

ENABLE: RAIN

AND OR DONE

ENABLE: RAIN AND

LEVEL FLOW TEMP5

pH5 SP_CO5

DO5 DONE

“RAIN” SET POINT:

.5 INCHES PER

__HOURS, __MINUTES

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

RESET RAIN HISTORY?

YES NO

“LEVEL” CONDITION

SET POINT

RANGE

RATE OF CHANGE

“LEVEL” SET POINT:

+1 ft

( 0.001 - 30.00)

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

ENABLE:

RAIN >0.50” / 2:00

AND

LEVEL >1.000 ft

Table 5-4 Sampler Enable

In this example, you will be enabling the sampler if it rains .5” in 2 hours and the level goes above 1 foot.

Select RAIN.

Choose AND.

Select LEVEL.

Enter .5 for the inches of rain. Press

.

Enter 2 HOURS.

Select ABOVE SET POINT.

Select YES.

Select SET POINT.

Enter 1 foot.

Select ABOVE SET POINT.

Your enable conditions will be summarized on the display. If correct, press to continue.

If necessary, make changes to the enable conditions.

5-24