Table 8-1 Setting the Clock and Calendar

6712FR Refrigerated Sampler Section 8 Maintenance

1

Table 8-1 Setting the Clock and Calendar

RUN

Select OTHER FUNCTIONS.

PROGRAM

 

VIEW REPORT

 

OTHER FUNCTIONS

 

 

 

2

3

4

5

MAINTENANCE

MANUAL FUNCTIONS PROGRAMMING STYLE

SET CLOCK

PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

ENTER TIME AND DATE:

HH:MM DD-MON-YY

14:00 22-JAN-01

SET CLOCK

PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

Select MAINTENANCE.

Select SET CLOCK.

Enter the time and date.

Press STOP to return to the main menu.

8.4.2 Pump Tube Alarm The sampler tracks the pump counts in both the forward and reverse cycles with a resettable counter. When the counter reaches the default count of 1,000,000 or the count you have entered, the sampler displays a message, “WARNING! REPLACE PUMP TUBE.” The message appears each time you turn the sampler on or run a program until you reset the counter.

After replacing the pump tube (see page 8-9), reset the count to zero so the sampler can begin tallying the pump counts for the new tube. Replacing the pump tube does not reset the counter.

Manual background WARNING

Pump may actuate without warning. To avoid injury, sampler must be off when pump housing cover is removed for inspection or tubing replacement.

Experience may suggest a significantly different pump tube life. You can change the alarm count to represent the pump tube life for your application. The example below shows how to set the alarm count and reset the pump counts.