Teledyne manual 6712FR Refrigerated Sampler Standard Programming

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 60
Image 60
Manual background

6712FR Refrigerated Sampler

Section 4 Standard Programming

Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample (Contin-

4

5

6

NUMBER OF BOTTLES:

1 2 4 8 12 24

BOTTLE VOLUME IS 1000 ml (300-30000)

SUCTION LINE LENGTH

IS 10 ft

(3-99)

Select 24.

Enter 1000.

Enter 7.

7

8

9

10

11

12

TIME PACED

FLOW PACED

FLOW BETWEEN SAMPLE EVENTS: 2 PULSES (1-9999)

SEQUENTIAL

BOTTLES/SAMPLE

SAMPLES/BOTTLE

RUN CONTINUOUSLY?

YES NO

1 BOTTLES PER

SAMPLE EVENT (1-24)

NO DELAY TO START

DELAYED START

CLOCK TIME

WAIT FOR PHONE CALL

Select FLOW PACED.

Enter the number of pulses between sample events.

Select BOTTLES/SAMPLE.

For this example, select NO by pressing an arrow until the option blinks. Then, press . Selecting YES allows the program to run indefinitely by repeating the sample distribution. Continuous sampling assumes that filled bottles are replaced with empty bottles at regular service intervals.

Enter 2.

Select CLOCK TIME.

13

14

15

START FLOW COUNT AT:

00:00

SELECT DAYS:

SU MO TU WE TH FR SA

DONE

PROGRAMMING COMPLETE

RUN THIS PROGRAM

NOW?

YES NO

Enter 6, then enter 0.

This is the time at which the sampler will start the pacing countdown.

Select MO, WE, and FR.

Use the arrow keys and to select the days. When the correct days are blinking, select DONE and press ↵.

Select NO.

4-4

Page 60
Image 60
Teledyne manual 6712FR Refrigerated Sampler Standard Programming