6712FR Refrigerated Sampler

Section 5 Extended Programming

When the sampler detects that it is disabled at start time, it sus- pends the program until it is enabled. Once enabled, the sampler collects a sample only when programmed to take a sample at the enable time. The diagram in Figure 5-4shows the sampler’s response when enabled or disabled at the start time.

Sampler enabled at start time

Sampler disabled at start time by signal from pin F

A sample event at the start time always occurs for time-paced programs (except for random). For flow-paced programs, it occurs when the program requires a sample at start time.

A sample event at enable time occurs only when the program requires a sample at enable.

Figure 5-4 Start Time Diagram for Sampler Enable Settings