Teledyne manual 6712FR Refrigerated Sampler Installation/Preparation, 1 Mounting the Controller

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 32
Image 32
6712FR Refrigerated Sampler

6712FR Refrigerated Sampler

Section 2 Installation/Preparation

Multiple-Bottle Sampling

Composite Sampling

Mounting Rod

Mounting Rod

 

 

Distributor Shaft

1/4-20 Set Screw

Extension

 

 

 

Flat Washer

Flat Washer

Wing Nut

Wing Nut

Stop Arm

 

 

Distributor arm

 

End of discharge tube

Nut

Distributor arm

below neck of bottle.

 

 

 

Assembly

 

Figure 2-1 Mounting the Controller

2.1.3 Installing the

The distributor arm is adjustable so that you can position the

Distributor Arm and

discharge tube over the bottles of each kit. Each bottle kit also

Discharge Tube

uses a discharge tube of a specific length. Table 2-1lists the

 

correct discharge tube lengths for each bottle kit. If you are using

 

bulk tubing to cut your own tubes, cut the ends of the tube

 

square, not at an angle, and cut the length accurately.

 

The two-piece distributor arm has four positions, marked on the

 

underside with the letters A, B, C, and D. Adjust the arm to the

 

correct position for the bottle kit being used. Table 2-1shows the

 

correct position for each kit. Thread the discharge tube through

 

the spring and arm so that the tube protrudes from the arm 1/16

 

to 1/8 inch. If too much tube protrudes from the arm it can catch

 

on bottles and jam the distributor arm.

 

Attach the arm to the distributor shaft extension. Secure the arm

 

to the extension with the black plastic nut. Slide the free end of

Distributor Arm

the discharge tube onto the bulkhead fitting. After installing the

distributor arm and discharge tube, run the distributor to each

 

 

bottle position to ensure proper installation.