5.16Sampling Reports

6712FR Refrigerated Sampler Section 5 Extended Programming

5.16Sampling Reports

5.17Other Functions

Other Functions Menu

RUN “PROGRAM NAME

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

• Select OTHER FUNCTIONS.

MAINTENANCE

MANUAL FUNCTIONS

SOFTWARE OPTIONS

HARDWARE

Adjust Suction Head – If you have chosen to manually enter the suction head, and are in the manual paused state, the scrolling menu will have an option to adjust the suction head. If you make changes to the suction head value, the sampler logs a changed suction head event.

The 6712FR records a variety of data while running a sampling program. Sample reports can be found in Section 4.

Section 4, Table 4-3contains a list of source codes and program events that may be recorded when using either standard or extended programming.

In extended programming, the “Other Functions” menu screen includes options for:

Maintenance (see Section 7)

Manual Functions (see Section 4)

Software Options.

Hardware

5.18 Software Options

The software option screens allow you to set up a number of dif- ferent software options:

Liquid Detector Enable

Display Backlighting

Normal/Quick View Screens

Event Marks

Measurement Interval

Pump Purge Counts

Dual Sampler Mode

Serial Output

Full Bottle Detection

Remote Access

 

 

Interrogator Connect Power

5.18.1Liquid Detector Enable/Disable

USE LIQUID DETECTOR?

YES NO

Under some conditions (pressurized lines, sampling foam), a more repeatable sample volume may be delivered by disabling the liquid detector. Selecting NO will turn the Liquid Detector off.

If the Liquid Detector is off:

you must enter the SUCTION HEAD

no RINSES or RETRIES will be available

you will not be able to CALIBRATE SAMPLE volumes

the RESULTS report will read a COUNT TO LIQUID of 0

A complete purge will occur between samples when using multiple bottles-per-sample distribution.