6712FR Refrigerated Sampler

Section 8 Maintenance

Table 8-2 Resetting the Pump Counter

1

2

3

4

5

6

7

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

MAINTENANCE

MANUAL FUNCTIONS PROGRAMMING STYLE HARDWARE

SET CLOCK

PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

1024000 PUMP COUNTS,

WARNING AT 1000000

1024000 PUMP COUNTS, RESET PUMP COUNTER?

YES NO

WARNING AT 1000000

PUMP COUNTS

(1 - 99)00000)

SET CLOCK

PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

Select OTHER FUNCTIONS.

Select MAINTENANCE.

Select PUMP TUBE ALARM.

The sampler displays this screen briefly to give you the current pump-count information. The first line shows you the pump counts tallied since the counter was last reset. The second line shows the current alarm setting.

To reset the counter to zero, select YES. Always reset the counter after replacing a pump tube. Select NO when merely checking the current count.

If necessary, change the pump-count alarm setting by typing the first two digits of the new setting. The sampler accepts entries between 1 and 99. For example, to increase the count to 1,500,000, enter 15.

Press STOP to return to the main menu.

8.4.3 Internal Battery

The internal lithium battery preserves the sampler’s memory

Warning

when power is removed. For an explanation of how to replace the

 

lithium battery, go to Section 8.8

 

.

Table 8-3 Checking and Replacing the Internal Battery

1

2

3

4

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

MAINTENANCE

MANUAL FUNCTIONS PROGRAMMING STYLE HARDWARE

SET CLOCK

PUMP TUBE ALARM

INTERNAL BATTERY

DIAGNOSTICS

WARNING!

REPLACE INTERNAL BATTERY BEFORE 17-OCT-97

Select OTHER FUNCTIONS.

Select MAINTENANCE.

Select INTERNAL BATTERY.

The sampler displays this screen briefly to show you the expiration date.

8-6

Page 158
Image 158
Teledyne 6712FR manual Resetting the Pump Counter, Checking and Replacing the Internal Battery