Teledyne 6712FR Refrigerated Sampler Standard Programming, Source Codes, Error Codes, 4-22

Models: 6712FR

1 244
Download 244 pages 57.15 Kb
Page 78
Image 78
Source Codes

6712FR Refrigerated Sampler

Section 4 Standard Programming

Table 4-3 Reports: Program Events, Source Codes, and Error Codes (Continued)

(The Source and Error Codes Listed below refer to why a sample was taken and what happened while it was being taken.)

 

 

Source Codes

*C

Calibrate sample

Sample volume delivered during calibration.

 

 

 

*D

Disable

Sample event at disable time.

 

 

 

*E

Enable

Sample event at enable time.

 

 

 

Ev

Event

Event-paced sample event.

 

 

 

*F

Flow

Flow-paced sample event.

 

 

 

*G

Grab sample

Sample volume delivered as a grab sample.

 

 

 

M

Command driven sample

A command driven sample.

 

 

 

PH

Phone command

Phone command sample.

 

 

 

R

Resume

Sample event at resume time.

 

 

 

*S

Start

Sample event at start time.

 

 

 

Sw

Switch time

Sample event at switch time.

 

 

 

TO

Time Override

Sample event at bottle-switch time.

 

 

 

*T

time

Time-paced sample event.

 

 

 

 

 

Error Codes

*DJ

Distributor Jammed

The distributor jammed.

 

 

 

*IP

sample in progress

Sample in progress when report was printing.

 

 

 

*L

pump latch open

Unable to take sample because the pump band was opened.

 

 

 

*ND

no distributor arm

There is no distributor arm or the stop is damaged.

 

 

 

*NL

No Liquid Detected

The sampler was unable to detect liquid.

 

 

 

LF

Not enough flow

Not enough flow for the minimum sample volume entered for samples with volumes

 

 

proportional to flow.

 

 

 

*NM

No More Liquid

After the sampler detected liquid and while the sample was being taken, the liquid

 

 

detector stopped detecting liquid.

 

 

 

M

Bottle Full

The composite bottle is full.

 

 

 

*O

Sampler Shut Off

The sampler was turned off with the On-Off key during the sample event.

 

 

 

Ov

Probable Overflow

The sampler was directed to take a sample that would not fit into the bottle. Occurs

 

 

for flow paced, time override programs only.

 

 

 

*P

Power Failed

The power supply failed during the sample event. Note: Power failures during five

 

 

consecutive sample events results in a LOW BATTERY shut-down.

 

 

 

*PJ

Pump Jammed

The pump jammed.

 

 

 

*Sk

Sample Skipped

Sample not taken.

 

 

 

*US

User Stopped

You pressed the Stop key during the sample event.

 

 

 

 

 

Figures 4-5,4-6,4-7,4-8,and 4-9show examples of reports con-

 

 

figured to show program settings, sampling results, combined

 

 

results, and a summary report.