4.13 Operating the Pump

6712FR Refrigerated Sampler Section 4 Standard Programming

4.13 Operating the Pump

Operating the Pump

GRAB SAMPLE

CALIBRATE VOLUME

OPERATE PUMP

MOVE DISTRIBUTOR

SELECT DIRECTION

PUMP FORWARD

PUMP REVERSE

PURGING. . .

PRESS THE RED ‘STOP’

KEY WHEN DONE

PURGED _____COUNTS

You can manually operate the sampler’s pump in the forward or reverse direction. You can do this to test the pump, draw a large sample, or purge the suction line.

When you run the pump manually, the sampler reports the number of pump counts during the pump’s operation. Use this feature to obtain an estimate of the pump counts required to purge a nonstandard suction line.

You can also manually run the pump from the main menu. Press the “1” key and then ↵ to run the pump in reverse. Press the “3” key and then ↵ to run the pump forward.

Operating the Pump CAUTION

Tests indicate that the pump produces sound levels in excess of 85db at one meter. PROLONGED EXPOSURE TO THIS SOUND LEVEL COULD RESULT IN HEARING LOSS AND REQUIRES THE USE OF PROTECTIVE EAR PLUGS.

4.14Moving the Distributor Arm

Moving the Distributor

Arm

GRAB SAMPLE

CALIBRATE VOLUME

OPERATE PUMP

MOVE DISTRIBUTOR

Select MOVE DISTRIBUTOR

GO TO BOTTLE ____

(1-MAX)

Enter the bottle number to which you want to move the distributor

NOW AT BOTTLE ____

The sampler lets you reposition the distributor arm. Use this feature to verify a bottle location when installing a bottle kit. You cannot move the distributor arm while the sampler is running a program.

4.14Moving the Distributor Arm CAUTION

DO NOT rotate the distributor arm manually; this will severely damage the distributor drive gears.

4-17