4.6 Start Times

6712FR Refrigerated Sampler

Section 4 Standard Programming

4.6 Start Times

It is important to understand the difference between the time at

 

which you run a program and the program’s start time. Running

 

a program simply means selecting RUN from the sampler’s main

 

menu. The start time is the time at which the sampler begins the

 

program’s first sample interval countdown. The start time is con-

 

trolled by your selections from the start time screens.

 

Each program contains start-time settings that tell the sampler

 

when to begin the program. When programming the sampler, you

 

can select one of four start time options: NO DELAY TO START,

 

DELAYED START, CLOCK TIME, or WAIT FOR PHONE CALL.

Select NO DELAY TO START when you want the sampler to start as soon as you select RUN.

Select DELAYED START when you want the sampler to delay from 1 to 999 minutes before starting the program.

Select CLOCK TIME when you want the sampler to begin the program at a specific time on at least one day of the week.

Select WAIT FOR PHONE CALL when you want the sampler to begin the program after receiving a remote command to start.

4.6.1How Do Start Times As soon as you select RUN from the main menu, the sampler

Work?checks the program’s start time settings. With a “DELAYED

START” setting, the sampler starts a countdown to the start time. The period between the time you select RUN and the start time you’ve specified in the program is called the “delay to start time.”

When you run a program with “CLOCK TIME” settings, the sampler also starts a countdown to start time until the day of week and clock times are met.

The sampler may not collect a sample at the start time. The sampler must be enabled, and the pacing selected must call for a sample at start. In any event, the sampler checks the pacing set- tings and begins the pacing-interval countdown.