5.19.7 Refrigerator

6712FR Refrigerated Sampler

Section 5 Extended Programming

• an external digital modem

recommended: Teledyne Isco’s CDMA cellular modem part # 60-5314-489

or GSM cellular modem part # 60-5314-563

Temperature5.19.8 Analog OutputManual backgroundManual backgroundManual background Note

If an internal voice modem is installed, the sampler will use that modem, and text messaging will be disabled.

 

When you enter each of the (up to three) phone numbers for the

 

sampler to call, if text messaging is active, a request for the TAP

 

(Telocator Alphanumeric Protocol) server number and communi-

 

cations parameters will be made.

 

As with voice messaging, the sampler will dial each telephone

 

number up to five times per alarm condition, until the message is

 

acknowledged.

5.19.7 Refrigerator

Teledyne Isco offers a temperature sensor intended for use with

Temperature

the 6712FR Refrigerated Sampler. The optional sensor connects

 

to the Rain Gauge port and uses I/O3 to transmit its temperature

 

readings to the 6712.

Manual backgroundManual backgroundManual backgroundManual backgroundManual background Note

If alarms are being used, the temperature sensor is simply plugged into the panel mount on the one- or two-alarm output module. Because the temperature sensor uses I/O3, it cannot be used with a three-alarm output module.

 

To configure an attached temperature sensor, select “FR-TEMP”

 

as the input for I/O3. The 6712 will display the NEW

 

HARDWARE SETUP screen. Select DONE and the 6712 recon-

 

figures its memory partitions to record the temperature

 

readings.

 

Temperature readings are updated approximately every 2.5

 

seconds. The readings are stored as FR-TEMP readings at the

 

data storage interval. You can retrieve a history of temperature

 

readings through the 6712’s reports. The temperature data also

 

can be used as a sampler enable condition.

5.19.8 Analog Output

Teledyne Isco offers three analog outputs as a factory installed

 

option. The option is installed inside the controller and includes a

 

six-pin connector to interface with external devices.

 

The analog outputs can be programmed for either 0-to-20 mA or

 

4-to-20 mA operation. With the exception of rain, any parameter

 

that the 6712 sampler is currently measuring can be used to

 

control any (or all) of the analog output channels. Programming

 

the outputs requires parameter levels at each end of the analog

 

range and then assumes a linear relationship between those

 

points. MANUAL CONTROL of each output is also available.