Emerson AC8 user manual Log In and Choose Setup System, Setup System Menu, Select Zone Menu

Models: AC8

1 128
Download 128 pages 38.48 Kb
Page 76
Image 76
Select Zone Menu

System and Control Options

7.10Setup Operation

The AC8’s operation features allow you to set up a rotation sequence to alternate which devices are operating and which are placed on Standby, as well as test devices while in Standby mode and specify what to do when a Standby device goes into alarm as it is powered up. You may specify delay times after an output changes state, directing the AC8 to wait before acknowledging inputs (hold delay time) or turn- ing an output on or off (restart delay time). With staging, you may con- figure the AC8 to turn a standby device on or off when an analog sensor detects a condition above or below specified levels.

To change any operation feature:

Log In and Choose Setup System

• From the Main Menu, use the arrows ↑↓ to choose System and Control, then press Enter ↵ (see 7.1 - Login for help).

• Enter your password at the Login screen.

• From the System and Control Menu, use the arrows ↑↓ to choose Setup System, then press Enter ↵ .

Setup System Menu

SETUP INPUTS SETUP COMMON ALARM SETUP SENSOR

↑↓ =NEXT =SELECT

SETUP ZONES SETUP OUTPUTS SETUP MODEM&PAGERS SETUP SYSTEM INFO

>SETUP OPERATION SETUP I/O MATRIX

Select Zone Menu

SELECT ZONE FOR

THE OPERATION

> ZONE 1

↑↓ =NEXT =EDIT

Select Setup Operation

• From the Setup System Menu, shown at right, use the arrows ↑↓ to choose Setup Operation and press Enter ↵ .

Select a Zone to Configure

• From the Select Zone Menu, use the arrows ↑↓ to choose the zone you want—for example, Zone_1—and press Enter ↵ .

Use the steps following Table 28 to access any of the Setup Opera- tion options.

Setup Operation Menu

> AUTO SEQUENCING OFF

STANDBY TESTING OFF FAILED STANDBY N ↑↓ =NEXT =SELECT

HOLD DELAY

: 10:00

RESTART

: 00:06

STAGING

: OFF

Table 28 shows the default settings for operation options: automatic sequencing, standby testing, failed standby, hold delay, restart and staging.

Table 28 Default settings - operation features

Feature

Default

Other Options

 

 

 

AUTO SEQUENCING

OFF (Deactivated)

ON (Activated), 12h (Activated for 12-hour intervals)

STANDBY TESTING

OFF (Deactivated)

ON (Activated)

FAILED STANDBY

N (Deactivated)

Y (Activated)

HOLD DELAY

10:00 (No delay)

Any time (in minutes and seconds) from 00:00 to 99:59

RESTART

00:06 (No delay)

Any time (in minutes and seconds) from 00:00 to 99:59

STAGING

OFF (Deactivated)

ON (Activated)

Figure 12 provides an overview of the Setup Operation screens available via the LCD interface.

Figure 12 Menu overview - Setup Operation menu

Setup Operation

AUTO SEQUENCING OFF

STANDBY TESTING OFF

FAILED STANDBY N

HOLD DELAY

: 10:00

RESTART

: 00:06

STAGING

: OFF

Setup Operation options

Auto Sequencing

AUTO SEQUENCING ON EVERY 03 DAYS

AT 00:00

See Section 7.10.1

Standby Testing

>STANDBY TESTING ON EVERY 03 DAYS

AT 00:00 FOR 10 MIN

See Section 7.10.2

Failed Standby

FAILED STANDBY N

See Section 7.10.3

Hold Delay

HOLD DELAY

10 MIN : 00 SEC

See Section 7.10.4

Restart

RESTART DELAY

00 MIN : 06 SEC

See Section 7.10.5

Staging

STAGING: ON

TARGET: +0070.0

SENSITIVITY: +005.0

See Section 7.10.6

68

Page 76
Image 76
Emerson AC8 user manual Log In and Choose Setup System, Setup System Menu, Select Zone Menu, Select Setup Operation