Croma24 Service Manual

Damage may occur to the micrometer gauge if the Carrier is moved without lifting up on the measuring tip. This action could also take the micrometer out of alignment and foul the results of the alignment.

7.While lifting up the measuring tip of the micrometer, slide the Carrier to the center of the stabilizer as shown in Figure 3-20. Position it as close to the screw as possible and drop the measuring tip onto the platen. Do this a couple of times to ensure an accurate reading.

Figure 3-20. Carrier Positions for Head Height Adjustment.

8.Move the left and right ends of the stabilizer bracket until a reading of 0.075 +/- 0.003” is observed. Read only the RED numbers on the micrometer gauge. The measurement of 0.075 +/-0.003” is used because the test cartridge being used does not have a print head attached. A 0.010” difference had to be added to compensate for the lack of a print head.

9.Tighten the screw on the center of the stabilizer.