Damage may occur to the micrometer gauge if the Carrier is moved without lifting up on the measuring tip. This action could also take the micrometer out of alignment and foul the results of the alignment.
7.While lifting up the measuring tip of the micrometer, slide the Carrier to the center of the stabilizer as shown in Figure
Figure 3-20. Carrier Positions for Head Height Adjustment.
8.Move the left and right ends of the stabilizer bracket until a reading of 0.075 +/- 0.003” is observed. Read only the RED numbers on the micrometer gauge. The measurement of 0.075
9.Tighten the screw on the center of the stabilizer.
MAINTENANCE