| Croma24 Service Manual | |
|
|
| ||
| Table | |||
|
|
|
|
|
| Symptoms | Possible cause | Solution |
|
|
| • flex cable | 1) perform Flex | |
|
| contacts dirty or | Cable Contact | |
|
| damaged | Cleaning | |
|
|
| 2) replace carrier | |
|
|
| assembly | |
|
| • carrier belt is | reinstall, check | |
|
| loose, too tight, | tension assembly, | |
|
| worn, or damaged | and/or replace belt | |
|
| • carrier bushings | replace bushings | |
|
| worn or damaged |
|
|
Fan Does Not
Power Up
• MPCB has | 1) reseat connection |
24 VDC at J3 | at MPCB to fan |
pins | 2) replace fan |
| |
• power not being | replace MPCB |
applied to fan |
|
SHOOTING | TROUBLE- |
|
|
Troubleshooting 85