Croma24 Service Manual

10.While lifting up the measuring tip of the micrometer, slide the Carrier to the left of the stabilizer as shown in Figure 3-

20.Position it as close to the screw as possible and drop the measuring tip onto the platen. Do this a couple of times to ensure an accurate reading.

11.Move the left end of the stabilizer bracket until a reading of 0.075” +/- 0.003” is observed. Read only the RED numbers on the micrometer gauge.

12.Tighten the screw on the left end of the stabilizer.

13.While lifting up the measuring tip of the micrometer, slide the Carrier to the right of the stabilizer as shown in Figure 3-

20.Position it as close to the screw as possible and drop the measuring tip onto the platen. Do this a couple of times to ensure an accurate reading.

14.Move the right end of the stabilizer bracket until a reading of 0.075” +/- 0.003” is observed. Read only the RED numbers on the micrometer gauge.

15.Tighten the screw on the right end of the stabilizer.

16.Repeat these steps as many times as necessary to ensure an accurate alignment.

64 Maintenance

Page 64
Image 64
Integra CROMA24 service manual Maintenance