Croma24 Service Manual

Table 4-1. Troubleshooting Table (cont).

Symptoms

Possible cause

Solution

No Power (cont)

• all DC output

1) check fuse on

 

voltages not

power supply

 

present (see

 

 

Figure 3-18 for

 

 

voltages)

2) replace power

 

 

supply

 

• all DC voltages

replace MPCB

 

present at MPCB

 

Media Does Not

 

 

Move

• perform Stepper

replace stepper

 

 

Motor Winding

motor

 

Resistance check

 

 

• rough motion

bad bearings -

 

while spinning

replace stepper

 

stepper motor

motor

 

• pinch rollers

adjust pinch rollers

 

not adjusted

 

 

correctly

 

 

• paper sensor

replace paper

 

not responding

sensor

 

• media control

Croma24 Control

 

switches on

Panel corrupted -

 

printer are

reinstall Control

 

operating correctly

Panel

• bad MPCB

replace MPCB

80 Troubleshooting

Page 80
Image 80
Integra CROMA24 service manual Media Does Not Move