Table 4-1. Troubleshooting Table (cont).

Symptoms

Possible cause

Solution

Carrier Axis

Failure

• dirty (or

perform Slide Shaft

lubricated) slide

Cleaning procedure

shaft

 

• perform Servo

replace servo motor

Motor Winding

 

Resistance check

 

• check servo

bad bearings -

motor for smooth

replace servo motor

movement

 

• obstruction in

remove obstruction

path of carrier

 

(may or may not

 

be visible)

 

• dirty encoder

perform Encoder

strip

Strip Cleaning

 

procedure

• damaged

replace encoder

encoder strip

strip

• bad encoder

replace encoder

sensor

sensor

• worn carrier

replace carrier

bushings

bushings

• loose trailing

reseat trailing cable

cable connections

connections at the

 

MPCB and the

 

carrier assembly