Connecting/Disconnecting Patient Cables

Disconnecting the Pads Patient Cable

To disconnect the pads patient cable:

1.Rotate the green locking mechanism on the patient cable in the direction (clockwise), of the blue arrow on the defibrillator until it stops (as shown in Figure 10-2).

2.Hold the locking mechanism in this position as you pull the patient cable away from the defibrillator.

Figure 10-2 Disconnecting the Pads Patient Cable

Blue arrow

10

M3500B HeartStart XLT Defibrillator/Monitor

10-3