ERC Protocol

If CPR is needed, press 3$86( . While paused, a timer indicates the elapsed time and the total duration of the Pause state, as shown:

Figure 3-15

Paused

Elapsed Time

Total Pause Duration

Shocks: 3

Timer: 35 / 60 00:01:40

RESUME ANALYZE

The total pause duration depends on the event preceding the Pause state. If

you ￿entered the Pause state:3

at the completion of a shock series or shortly after a shock is deliv- ered, the duration is equal to the 3RVW￿6KRFN￿&35￿7LPHU configuration set-

￿ting (the default setting is 60 seconds).

when no shock was advised, the duration is equal to the ￿16$￿￿7LPHU configuration setting, where NSA is an acronym for No Shock Advised (the default setting is 180 seconds).

M3500B HeartStart XLT Defibrillator/Monitor

3-17