Defibrillating

3. If shock advised, press 6+2&. .

Once charging is complete, the charge tone becomes continuous. Make sure no one is touching the patient or anything connected to the patient. Call out "Clear." Then press 6+2&. to deliver a shock to the patient.

Figure 3-8 Press SHOCK Display

 

 

Stand Clear!

150J

 

 

Press SHOCK

 

 

 

Shocks: 0

 

 

 

00:00:14

 

 

 

SHOCK

 

 

 

WARNING

 

Defibrillation current can cause operator or bystander injury. Do not touch the

 

 

patient, or equipment connected to the patient, during defibrillation.

 

 

 

 

 

The defibrillator automatically disarms if you do not press 6+2&. in 30 seconds.

3-10

Defibrillating in AED Mode

Page 44
Image 44
Philips M3500B manual If shock advised, press 6+2, 150J, Stand Clear, Press Shock