Selecting a Sensor

 

 

 

 

 

 

 

 

 

 

 

 

Selecting a Sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 5-1 shows the SpO2 sensors that may be used with the HeartStart XLT.

 

 

 

Table 5-1 Approved Sensors

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensor

Type

Patient

Patient Size

 

Ideal Site

 

 

 

 

 

 

 

 

 

 

 

 

M1191A

Reusable

Adult

> 50 kg

 

Finger

 

 

 

 

 

 

 

 

 

 

 

 

M1192A

Reusable

Small adult

15-50 kg

 

Finger

 

 

 

 

 

Pediatric

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M1194A

Reusable

Pediatric

> 40 kg

 

Fleshy part

 

 

 

 

 

Adult

 

 

of ear

 

 

 

 

 

 

 

 

 

 

 

 

M1903A/B

Disposable

Pediatric

10-50 kg

 

Toe/Finger

 

 

 

(Nellcor D-20)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M1904A/B

Disposable

Adult

> 30 kg

 

Finger

 

 

 

(Nellcor D-25)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M1906A

Semi-reus-

Pediatric

3-40 kg

 

Finger/Toe

 

 

 

(Nellcor P/I)

able

Infant

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M1907A

Semi-reus-

Neonate

< 3 kg

 

Foot/Hand

 

 

 

(Nellcor A/N)

able

 

 

 

 

 

 

 

Adult

> 40 kg

 

Finger

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE

 

 

To use Nellcor sensors, you must connect the M1943A Nellcor Adaptor

 

 

 

patient cable to the HeartStart XLT. (See “Connecting the SpO2 Patient

 

 

 

 

 

 

Cable” on page 10-5.)

5

M3500B HeartStart XLT Defibrillator/Monitor

5-3

Page 63
Image 63
Philips M3500B manual Selecting a Sensor, Approved Sensors, Sensor Type Patient Patient Size Ideal Site