Specifications
M3500B HeartStart XLT Defibrillator/Monitor 13-9
13
Environmental
Temperature: 0o to 50oC operating, -20o to 70oC storage.
Thermal paper may darken above 55oC.
Charging the battery at temperatures above 35oC may degrade batt ery
life.
Storing the battery for extended periods at temperatures above 40oC
will reduce battery capacity and degrade battery life.
Operating and storage specifications for electrodes may vary. Refer to
the manufacturer’s specifications for details.
Humidity:
Up to 95% Relative Humidity
Printer paper may jam if paper is wet.
Thermal printer may be damaged if wet paper is allowed to dry while
in contact with printer elements.
Altitude:
Operating: up to 15,000 ft.
Shipping: up to 20,000 ft.
Shock (drops onto concrete): Unit survives 39" (1 m) drops onto all surfaces
(faces, corners, and edges) enclosed in carrying case. Exposed handle surface s
survive 24" drops.
Vibration: Mil Std 810E 514.4 Category 6 Helicopter, General Storage,
UH60.
Water Resistance: IPX4, splash proof per IEC 60529, with Data Card door
closed.
EMC: Meets EN 60601-1-2. This ISM device complies with Canadian
ICES-001.
Other Considerations: Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: Continuous.