Specifications

Environmental

Temperature: 0o to 50oC operating, -20oto 70oC storage.

Thermal paper may darken above 55oC.

Charging the battery at temperatures above 35oC may degrade battery life.

Storing the battery for extended periods at temperatures above 40oC

￿will reduce battery capacity and degrade battery life.

Operating and storage specifications for electrodes may vary. Refer to the manufacturer’s specifications for details.

Humidity:

Up to 95% Relative Humidity

￿Printer paper may jam if paper is wet.

Thermal printer may be damaged if wet paper is allowed to dry while in contact with printer elements.

Altitude:

13

Operating: up to 15,000 ft.

Shipping: up to 20,000 ft.

Shock (drops onto concrete): Unit survives 39" (1 m) drops onto all surfaces (faces, corners, and edges) enclosed in carrying case. Exposed handle surfaces survive 24" drops.

Vibration: Mil Std 810E 514.4 Category 6 Helicopter, General Storage, UH60.

Water Resistance: IPX4, splash proof per IEC 60529, with Data Card door closed.

EMC: Meets EN 60601-1-2. This ISM device complies with Canadian ICES-001.

Other Considerations: Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.

Mode of Operation: Continuous.