Specifications

Noninvasive Pacing

Waveform: Monophasic Truncated Exponential.

Current Pulse Amplitude: 10 mA to 200 mA (5 mA resolution); accuracy 10 mA - 50 mA + 5 mA, 50 mA - 200 mA + 10%.

Pulse Width: 20 ms with accuracy +0, -5 ms.

Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy + 1.5%.

Modes: Demand or Fixed Rate.

Refractory Period: 340 msec (40 to 80 ppm); 240 msec (90 to 180 ppm).

SpO2/Pulse Oximetry

Accuracy with:

M1191A sensor - 1 standard deviation 70% to 100%, +

2.5%.

 

M1192A sensor - 1 standard deviation 70% to 100%, +

2.5%.

 

M1194A sensor - 1 standard deviation 70% to 100%, +

4.0%.

13

NELLCOR sensors - 1 standard deviation 80% to 100% + 3.0%. Resolution: 1%.

SpO2 Alarm Limits: Three preset low alarm limits: 90, 85, and 80%, default is off.

INOP Alerts: Triggered by disconnected sensor, noisy signal, light interfer- ence or low signal (non-pulsatile).

M3500B HeartStart XLT Defibrillator/Monitor

13-7

Page 153
Image 153
Philips M3500B manual Noninvasive Pacing, SpO2/Pulse Oximetry